Gerätetyp
|
Pressure Transducer Display |
Power Supply, Readout & Setpoint Controller |
30 Channel Controller, Power Supply for up to 10 MFCs |
Unterscheidungsmerkmal
|
Miniature display for point-of-use pressure monitoring |
Innovative solution for up to four mass flow and/or pressure devices |
Turnkey solution for up 30 MFCs |
Power Input
|
4-20 mA (2-wire, loop powered) |
Voltage: 12-24 Vdc required, -15 Vdc permitted Current: 400 mA max current draw per channel Instrument power draw: 0.8 Watts Optional power module: 100-240 Vac, 47-63 Hz |
85-250 Vac, 47-63 Hz |
Spannungsausgang
|
Current loop powered (transducer voltage +6 Vdc loop voltage drop) |
+15V/2.0A, -15V/1.0A or 12-24 Vdc/2.0A |
Voltage: 24 Vdc (± 10%) Current: 3.5 Amp Will power up to 10 Brooks S-Series, SLA Smart II or 4800 Series mass flow devices |
Signal Ein- und Ausgang
|
Input: Connector, factory wired for 4-20 mA transducer Output: Top mount, 6-wire cable |
0(1)-5 Volts, 0(2)-10 Volts, 0(4)-20 mA |
RS485 S-Protocol (HART Command Set) |
Digitale Kommunikation
|
— |
Full communications capability for remote readout, setpoint, control, programming and data acquisition via RS232. |
RS485 S-Protocol |
Montageoptionen
|
Top |
Panel mount kit
Table top kit
Rack mount kit (0254 only, can be used to retrofit into 0152/54 installations)
Rack mount kit with 19” rack (0254 only) |
Table top |
EMV
|
CE Mark EN61326 |
CE Mark: EMC Directive EN61326-1:2013
FCC Part 15 Class A, Part 68 |
CE Mark EN61326-1 |
Einhaltung von Umweltvorschriften
|
RoHS Directive (2011/65/EU)
REACH Directive EC 1907/2006 |
RoHS EPD 2011/65/EU
UL-USA & Canada UL 61010 electrical safety for general purpose indoor use |
RoHS EPD 2002/95/EC, 01Jul2006 |